VEGA MINITRAC 31 4 … 20 mA_HART - four-wire User Manual
Page 46

46
6 Set up with the display and adjustment module
MINITRAC 31 • 4 … 20 mA/HART - four-wire
40447-EN-131119
Menu
Menu item
Default value
Setup
Measurement loop name
Sensor
Isotope
Cs-137
Application
Density
Background radiation
0 ct/s
Unit of the process value
g/cm
3
Temperature unit
° C
Adjustment min. process
value
0.500 g/cm
3
Adjustment max. process
value
1.500 g/cm
3
Inner diameter
0.20 m
Linearization
90000 ct/s and 0.500 g/cm
3
Damping
60 s
Real value correction
0
Current output mode
4 … 20 mA, < 3.6 mA
Current output Min./Max.
Min. current 3.8 mA, max. current 20.5 mA
X-ray alarm
Modulated measuring current
Reference value - Relay
None
Block adjustment
Released
Display
Language
Selected language
Displayed value
Pulse rate
Display unit
ct/s
Additional adjustments
HART mode
Standard
Address 0
With this function you can select the mode.
The sensor offers the HART modes standard and multidrop.
The mode 'Standard' with the fixed address 0 (factory setting) means
output of the measured value as 4 … 20 mA signal.
This parameter is described in the operating instructions manual
"Display and adjustment module".
With this function
•
Load parameter adjustment data from the sensor into the display
and adjustment module
•
Write parameter adjustment data from the display and adjustment
module into the sensor
This parameter is described in the operating instructions manual
"Display and adjustment module".
Info
In this menu you will find the following menu items:
Additional adjustments/
HART mode
Additional adjustments -
Copy instrument settings
Info