36

6 Set up with the display and adjustment module

MINITRAC 31 • 4 … 20 mA/HART - four-wire

40447-EN-131119

The natural radiation on earth influences the accuracy of the meas-

urement.
With this menu item the natural background radiation can be faded

out.
Note:

Make sure that some products have a self-radiation. This is for ex-

ample the case with oil or potash salt lye. Therefore the tube must be

filled to determine the background radiation.
For this purpose, the MINITRAC 31 measures the natural background

radiation and sets the pulse rate to zero.
In the future, the pulse rate from this background radiation will be

automatically deducted from the total pulse rate. This means: only

the component of the pulse rate originating from the source will be

displayed.
The source container must be closed for this setting (OFF).

In this menu item you can select the units of the process value and

the temperature.

In this menu item you can enter the measuring range (min. and max.

process value) of the sensor.
These settings influence the current output of the sensor.

Enter in the menu window "Max. process value" the max. density

value, for example in "g/cm

3

". This corresponds to an output current

of 20 mA.

Enter in the menu window "Min. process value" the min. density value,

for example in "g/cm

3

". This corresponds to an output current of 4 mA.

Setup/Background radia-

tion

Setup - Units

Setup - Adjustment