8
.3 Exchange or shorten cable/rod
T
he cable or rod (meas. part) of the probe can be shortened, if
necessary. To loosen the meas. part you need two fork spanners with
spanner width 8.
1
L
oosen the measuring component by applying a fork spanner to
the flat surfaces (SW 8), provide counterforce with another fork
spanner (SW 8)
2
U
nscrew the loosened measuring component manually
3
P
lace the enclosed new double washer onto the thread.
Caution:
M
ake sure that the two components of the double washer remain
together.
4
S
crew in a new measuring part manually
5
E
xert counterforce with the second fork spanner and tighten the
measuring component on the flat surfaces with a torque of 15 Nm
(11
lbf ft).
Fig. 27: Exchanging the cable or rod
Information:
P
lease keep the stated torque so that the max. tensile strength
remains.
T
he cable and rod of the probe can be shortened by any amount.
1
C
able: loosen the three pins on the gravity weight (hexagon 4)
2
C
able: remove the pins
Exchanging the cable/
rod
Shorten cable/rod
VEGAFLEX
62 • 4 … 20
mA/HART two-wire