beautypg.com

3 function chart – VEGA VEGAVIB 67 - NAMUR User Manual

Page 21

background image

case of very light-weight solids, you have to turn the

potentiometer to the complete left position (>0.008 g/cm³ or

>0.0003 lbs/in³). Hence the VEGAVIB 67 will be more

sensitive and can detect light-weight solids more reliably.
For instruments detecting solids in water, these values are not

applicable. The potentiometer is preset and must not be

changed.

The characteristics reversal can be carried out with the DIL

switch. You can choose between falling characteristics (switch

position max.) and rising characteristics (switch position min.).

You can have the requested current outputted.
Modes

l

min. - rising characteristic curve (High current when

immersed)

l

max. - falling characteristics (Low current when immersed)

The NAMUR output can be switched over to falling or rising

characteristics (see also "Function chart").

The simulation key is located in a recess on the upper side of

the oscillator. Push the simulation key with a suitable object

(screwdriver, pen, etc.).
When the key is pushed, line break between sensor and

processing unit is simulated. The signal lamp on the sensor

extinguishes. The measuring system must signal failure and

take on safe condition when the key is pushed.
Keep in mind that the connected instruments are activated

during operation. By doing this, you can check the correct

function of the measuring system.

Control lamp (LED) for indication of the switching condition

l

yellow = High current >= 2.2 mA

l

dark = Low current <=1 mA

l

yellow (flashing) = Failure <= 1 mA

6.3 Function chart

Level switch VEGAVIB 67

The following chart provides an overview of the switching

conditions depending on the adjusted mode and level.

Characteristics reversal (2)

Simulation key (3)

Signal lamp (6)

VEGAVIB 67 - with NAMUR output

21

Set up

31220

-EN

-060508