background image

Power supply by/max. number of sensors

-

H

1 power supply

max. 32 (max. 10 with Ex)

Electrical protective measures

Protection, depending on housing version

-

Plastic housing

IP

66/IP 67

-

Aluminium housing, stainless steel
housing - investment casting, stainless
steel housing - electro-polished

IP

66/IP 68 (0.2 bar)

5)

-

Aluminium and stainless housing, in-
vestment casting (optionally available)

IP

66/IP 68 (1 bar)

Overvoltage category

III

Protection class

II

Approvals

Instruments with approvals can have different technical data depending on the version.

That's why the associated approval documents have to be noted with these instruments. They are
part of the delivery or can be downloaded under www.vega.com via "VEGA Tools" and "serial
number search
" as well as via "Downloads" and "Approvals".

5)

A suitable cable is the prerequisite for maintaining the protection rating.