6 setup, 1 general information, 2 adjustment elements – VEGA VEGACAP 69 - contactless electronic switch User Manual
Page 17
17
6 Setup
VEGACAP 69 • - contactless electronic switch
31174-EN-130916
6 Setup
6.1 General information
The figures in brackets refer to the following illustrations.
On the electronics module you will find the following display and
adjustment elements:
•
Potentiometer for switching point adaptation
•
DIL switch for measuring range selection
•
DIL switch for mode adjustment - min./max.
•
Control lamp
Note:
As a rule, always set the mode with the mode switch (3) before start-
ing setup VEGACAP 69. The switching output will change if you set
the mode switch (3) afterwards. This could possibly trigger other con-
nected instruments or devices.
6.2 Adjustment elements
4
5
6
3
2
1
Fig. 10: Oscillator - Contactless electronic switch
1 Potentiometer for switching point adaptation
2 DIL switch for measuring range selection (with compensation button)
3 DIL switch for mode adjustment
4 Ground terminal
5 Connection terminals
6 Control lamp
The switching status of the electronics can be checked with closed
housing (only plastic housing), see "Function chart".
Note:
Screw the housing cover tightly up to the thread stop so that the
inspection glass is above the control lamp (LED).
To adjust VEGACAP 69, first of all remove the housing cover.
Function/Configuration