beautypg.com

VEGA VEGAFLEX 63 4 … 20 mA_HART two-wire User Manual

Page 51

background image

Influence of the ambient temperature to the sensor electronics
T

emperature drift

0.03 %/10 K

relating to the max. measuring range

or max. 0.3 %

T

emperature drift - Digital output

3

mm/10 K relating to the max. measuring range or

max. 10 mm

Ambient conditions

A

mbient, storage and transport temperature

-40 … +80 °C (-40 … +176 °F)

Process conditions

P

rocess pressure

-

F

lange version ≤ 2"/DN 50

-0.5 … +16

bar/-50 … +1600 kPa (-7.3 … +232 psig),

depending on the process fitting

-

F

lange version > 2"/DN 50

-0.2 … +16

bar/-20 … +1600 kPa (-2.9 … +232 psig),

depending on the process fitting

P

rocess temperature (flange temperature)

-40 … +150 °C (-40 … +302 °F)

T

he measurement error from the process condi-

tions is in the specified pressure and temperature
range of below 1 %.

0 °C

(32 °F)

40 °C

(104 °F)

20 °C

(58 °F)

60 °C

(140 °F)

80 °C

(176 °F)

50 °C

(122 °F)

-40 °C

(-40 °F)

100 °C

(212 °F)

150 °C

(302 °F)

1

2

Fig. 29: Ambient temperature - Process temperature

1

Ambient temperature

2

Process temperature (depending on the seal material)

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar

C

able entry/plug

3)

-

S

ingle chamber housing

l

1

x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),

1

x blind stopper M20 x 1.5

or:

l

1

x closing cap M20 x 1.5; 1 x blind stopper

M

20

x 1.5

3)

D

epending on the version M12 x 1, according to DIN 43650, Harting, 7/

8" FF.

VEGAFLEX

63 • 4 … 20

mA/HART two-wire

51

10 S

upplement

31841

-

EN

-100426