3 exchanging the electronics module – VEGA VEGAFLEX 65 Profibus PA User Manual
Page 36
Status accor-
ding to NE 107
Error code
Text message
Cause/Rectification
E
017
Adjustment span too small
Adjustment not within the specifica-
tion. Carry out the adjustment again,
increasing the distance between
min. and max. adjustment
E
036
No operable software
Failed or interrupted software up-
date/Repeat software update
E
042
Hardware error, electronics defec-
tive
Exchange the instrument or send it
in for repair
E
043
Hardware error, electronics defec-
tive
Exchange the instrument or send it
in for repair
E
113
Communication conflict
Exchange the instrument or send it
in for repair
Depending on the reason for the fault and the measures taken, the
steps described in chapter "Set up" may have to be carried out again.
8.3 Exchanging the electronics module
If the electronics module is defective, it can be replaced by the user.
In Ex applications only one instrument and one electronics module with
respective Ex approval may be used.
If there is no electronics module available on site, one can be ordered
from the VEGA agency serving you.
The new electronics module must be loaded with the settings of the
sensor. These are the options:
l
At the factory by VEGA
l
Or on site by the user
In both cases, the sensor serial number is necessary. The serial
numbers are stated on the type label of the instrument, inside the
housing or on the delivery note.
Information:
When loading on site, first of all the order data must be downloaded
from the Internet (see operating instructions manual "Electronics
module").
The electronics modules are adapted to the respective sensor and
distinguish also in the signal output or power supply.
Electronics module FX-E.60P suitable for VEGAFLEX 61, 62, 63, 65,
66 - Profibus PA:
Reaction after fault rec-
tification
Sensor serial number
Assignment
Profibus PA
36
VEGAFLEX
65 • Profibus PA
8 Maintenance and fault rectification
31847
-EN
-120329