beautypg.com

VEGA VEGACAL 67 4 … 20 mA_HART - two-wire User Manual

Page 43

background image

43

10 Supplement

VEGACAL 67 • 4 … 20 mA/HART - two-wire

31758-EN-130917

Ʋ Double chamber housing

1␣x IP␣68 cable gland M20␣x␣1.5; 1␣x blind stopper

M20␣x␣1.5; 1␣x blind stopper M16␣x␣1.5

Connection cable

Ʋ Wire cross-section

0.5 mm² (AWG 20)

Ʋ Wire resistance

< 0.036 Ω/m (0.011 Ω/ft)

Ʋ Tensile strength

< 1200 N (270 lbf)

Ʋ Standard length

5 m (16.4 ft)

Ʋ Max. length

1000 m (3280 ft)

Ʋ Min. bending radius

25 mm (0.984 in) with 25 °C (77 °F)

Ʋ Diameter approx.

8 mm (0.315 in)

Ʋ Colour - standard PE

Black

Ʋ Colour - standard PUR

Blue

Ʋ Colour - Ex-version

Blue

Display and adjustment module
Voltage supply and data transmission

through the sensor

Indication

LC display in dot matrix

Adjustment elements

4 keys

Protection rating

Ʋ unassembled

IP 20

Ʋ mounted into the sensor without cover IP 40

Material

Ʋ Housing

ABS

Ʋ Inspection window

Polyester foil

Voltage supply
Operating voltage

Ʋ Non-Ex instrument

12 … 36 V DC

Ʋ EEx-ia instrument

12 … 30 V DC

Ʋ EEx-d-ia instrument

18 … 36 V DC