VEGA VEGAFLEX 65 Foundation Fieldbus User Manual

Page 40

background image

Transmission rate

31.25 Kbit/s

Current value

10 mA, ±0.5 mA

Resolution, digital

> 1

mm (0.039 in)

Accuracy (according to DIN EN 60770-1)

Process reference conditions according to DIN EN 61298-1

-

Temperature

+18

… +30 °C (+64 … +86 °F)

-

Relative humidity

45 … 75 %

-

Air pressure

+860

… +1060 mbar/+86 … +106 kPa

(+12.5 … +15.4 psig)

Installation reference conditions

-

Min. distance to installations

> 500

mm (19.69 in)

-

Vessel

metallic, ø 1 m (3.281 ft), centric installation,
process fitting flush with the vessel ceiling

-

Medium

Water/Oil (dielectric figure ~2.0)

-

Installation

Probe end does not touch the vessel bottom

Sensor parameter adjustment

Gating out of false signals carried out

Deviation

see diagrams
Depending on the installation conditions, there can
be deviations which can be rectified with an
adaptation of the adjustment or a change of the
measured value offset in the DTM service mode.

L

0

-2 mm (-0.079

")

2 mm (0.079

")

-10 mm (0.394

")

5 mm (0.197

")

0,15 m (5.906

"

)

0,03 m (1.181

"

)

0,05 m

(1.969

")

1

Fig. 17: Deviation VEGAFLEX 65 in coax version in water

1

Dead zone - no measurement possible in this area

L

Probe length

40

VEGAFLEX

65 • Foundation Fieldbus

10 Supplement

31848

-EN