l

FX-E.67PX (X = without approvals)

l

FX-E.67PC (C = approvals XM, CX, CM according to product list)

8.4 Software update

The software version of VEGAFLEX 67 can be determined as follows:

l

on the type label of the electronics

l

via the indicating and adjustment module

l

via PACTware

You can view all software histories on our website www.vega.com.
Make use of this advantage and get registered for update information
via e-mail.

The following components are required to update the sensor software:

l

Sensor

l

Voltage supply

l

VEGACONNECT

l

PC with PACTware

l

Current sensor software as file

At "www.vega.com/downloads" go to "Software". Select under "plics
sensors and instruments
", "Firmware updates" the respective instru-
ment series and software version. Load the zip file via the right mouse
key with "Save target as" e.g. on the desktop of your PC. Move with the
right mouse key to the folder and select "Extract all". Save the
extracted files, for example on the desktop.

Connect the signal conditioning instrument to power supply and
provide the connection from the PC to the instrument via the interface
adapter. Start PACTware and go via the menu "Project" to the VEGA
project assistant
. Select "USB" and "Set instruments online". Activate
the project assistant with "Start". The assistant establishes the
connection automatically and opens the parameter adjustment window
"S

ensor # online parameter adjustment". Connect this parameter

adjustment window before you carry out further steps.

Select with the right mouse key the sensor in the project and go to
"A

dditional function". Then click to "Software update". The window

"S

ensor # software update" opens. PACTware checks now the sensor

data and displays the actual hardware and software version of the
sensor. This takes approximately 60 s.

Push the button "Update software" and select the previously extracted
hex file. Then the software update can be started. The additional files
are installed automatically. Depending on the sensor, this procedure
lasts up to 1 h. Then the message appears ""Software update
successfully executed
".

8.5 Instrument repair

If a repair is necessary, please proceed as follows:

Load sensor software to
PC

Prepare update

Load software into sen-
sor

VEGAFLEX

67 • Profibus PA

39

8 Maintenance and fault rectification