beautypg.com

3 wiring plan, double chamber housing – VEGA VEGAFLEX 67 (-40…+150°C) 4 … 20 mA_HART four-wire User Manual

Page 17

background image

6

I

nsert the wire ends into the open terminals according to the wiring

plan

7

P

ress down the opening levers of the terminals, you will hear the

terminal spring closing

8

C

heck the hold of the wires in the terminals by lightly pulling on

them

9

C

onnect the screen to the internal ground terminal, connect the

outer ground terminal to potential equalisation

10 T

ighten the compression nut of the cable entry. The seal ring must

completely encircle the cable

11 C

onnect the lead cable for power supply in the same way

according to the wiring plan, in addition connect the ground
conductor to the inner ground terminal.

12 S

crew the housing cover back on

T

he electrical connection is finished.

Fig. 6: Connection steps 5 and 6

5

.3 Wiring plan, double chamber housing

T

he following illustration apply to non-Ex as well as Ex ia versions. The

E

xd version is described in the next subchapter.

VEGAFLEX

67 • 4 … 20

mA/HART - four-wire

17

5 C

onnecting to power supply

31854

-

EN

-120329