4 software update – VEGA VEGAFLEX 67 (-40…+150°C) 4 … 20 mA_HART two-wire User Manual
Page 41
I
n Ex applications only one instrument and one electronics module with
respective Ex approval may be used.
I
f there is no electronics module available on site, one can be ordered
from the VEGA agency serving you.
T
he new electronics module must be loaded with the settings of the
sensor. These are the options:
l
A
t the factory by VEGA
l
O
r on site by the user
I
n both cases, the sensor serial number is necessary. The serial
numbers are stated on the type label of the instrument, inside the
housing or on the delivery note.
Information:
W
hen loading on site, first of all the order data must be downloaded
from the Internet (see operating instructions manual "Electronics
module").
T
he electronics modules are adapted to the respective sensor and
distinguish also in the signal output or power supply.
E
lectronics module FX-E.67H suitable for VEGAFLEX 67 - 4 … 20 mA/
HART:
l
FX-E.67HX (X =
without approvals)
l
FX-E.67HC (C =
approvals XM, CX, CM, CK, CI, DX, DM, DI
according to product list)
8
.4 Software update
T
he software version of VEGAFLEX 67 can be determined as follows:
l
on the type label of the electronics
l
via the indicating and adjustment module
l
via PACTware
Y
ou can view all software histories on our website www.vega.com.
M
ake use of this advantage and get registered for update information
via e-mail.
T
he following components are required to update the sensor software:
l
S
ensor
l
V
oltage supply
l
VEGACONNECT
l
PC
with PACTware
l
C
urrent sensor software as file
Sensor serial number
Assignment
4
… 20 mA/HART
VEGAFLEX
67 • 4 … 20
mA/HART - two-wire
41
8 M
aintenance and fault rectification
31853
-
EN
-120329