VEGA VEGASON 61 Profibus PA User Manual
Page 42
42
10 Supplement
VEGASON 61 • Profibus PA
28784-EN-130417
Measuring characteristics
Ultrasonic frequency
70 kHz
Interval
> 2 s (dependent on the parameter adjustment)
Abstrahlwinkel at -3 dB
11°
Step response or adjustment time
5)
> 3 s (dependent on the parameter adjustment)
Measuring accuracy
Resolution, general
> 1 mm (0.039 in)
Deviation
6)
see diagram
5 m (16.404 ft)
4 m (13.123 ft)
3 m (9.843 ft)
2 m (6.562 ft)
1 m (3.28 ft)
10 mm (0.394 in)
4 mm (0.157 in)
-10 mm (-0.394 in)
-4 mm (-0.157 in)
Fig. 28: Deviation VEGASON 61
Influence of the ambient temperature to the sensor electronics
7)
Average temperature coefficient of the
zero signal (temperature error)
0.06 %/10 K
Ambient conditions
Ambient, storage and transport tempera-
ture
-40 … +80 °C (-40 … +176 °F)
Process conditions
Process pressure
-20 … 200 kPa/-0.2 … 2 bar (-2.9 … 29 psig)
Process temperature (transducer tem-
perature)
-40 … +80 °C (-40 … +176 °F)
Vibration resistance
mechanical vibrations with 4 g and 5 … 100 Hz
8)
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug
9)
5)
Time to output the correct level (with max. 10 % deviation) after a sudden level change.
6)
Incl. non-linearity, hysteresis and non-repeatability.
7)
Relating to the nominal measuring range.
8)
Tested according to the guidelines of German Lloyd, GL directive 2.
9)
Depending on the version M12 x 1, according to DIN 43650, Harting, 7/8" FF.