4 Mounting

VEGASON 62 • Profibus PA

28785-EN-130417

ca. 10 mm

Fig. 7: Recommended socket mounting

If the reflective properties of the medium are good, you can mount

VEGASON 62 on sockets which are higher than the length of the

transducer. You will find recommended values for socket heights in the

following illustration. The socket end should be smooth and burr-free,

if possible also rounded. Carry out a false echo storage.

d

d

h

100 mm/4"

150 mm/6"

200 mm/8"

300 mm/12"

400 mm/16"

500 mm/20"

h

Fig. 8: Deviating socket dimensions

Align the sensor in liquids as vertical as possible to the product sur-

face to achieve an optimum measurement result.

Fig. 9: Alignment in liquids

To reduce the min. distance to the medium, you can also mount VE-

GASON 62 with a beam deflector. By doing this, it is possible to fill the

vessel nearly to maximum. Such an arrangement is suitable primarily

for open vessels such as e.g. overflow basins.

Sensor orientation