4 Mounting

VEGASON 62 • Profibus PA

28785-EN-130417

When mounting the sensor, keep a distance of at least 200 mm

(7.874 in) to the vessel wall. If the sensor is installed in the center of

dished or round vessel tops, multiple echoes can arise. These can,

however, be suppressed by an appropriate adjustment (see chapter

"Setup").

1

2

> 200 mm

Fig. 5: Mounting on round vessel tops
1 Reference plane

2 Vessel center or symmetry axis

If you cannot keep this distance you should carry out a false echo

storage before setup. This applies mainly if buildup on the vessel wall

is expected. In this case, we recommend repeating a false echo stor-

age later with existing buildup.
In vessels with conical bottom it can be advantageous to mount the

sensor in the center of the vessel, as measurement is then possible

down to the lowest point of the vessel bottom.

Fig. 6: Vessel with conical bottom

Socket pieces should be dimensioned so that the lower end of the

transducer protrudes at least 10 mm (0.394 in) out of the socket.

Installation position

Socket