background image

?

4 … 20 mA signal missing

l

Wrong connection

à

Check connection according to chapter "Connection

procedure" and, if necessary, correct according to

chapter "Wiring plan"

l

No supply voltage

à

check cables for line break, repair, if necessary

l

supply voltage too low or load resistance too high

à

Check, adapt, if necessary

?

Current signal greater than 22 mA or less than 3.6 mA

l

Electronics defective

à

Exchange instrument or return instrument for repair

In Ex applications, the regulations for the wiring of intrinsically

safe circuits must be observed.

?

E013

l

no measured value available

à

sensor in boot phase

à

sensor does not find an echo, e.g. because of faulty

installation or incorrect parameter adjustment

?

E017

l

Adjustment span too small

à

Carry out a fresh adjustment and increase the distance

between min. and max. adjustment

?

E036

l

no operable sensor software

à

Carry out software update or send instrument for repair

?

E041

l

Hardware error, electronics defective

à

Exchange instrument or return instrument for repair

Fault messages via the indi-

cating/adjustment module

Maintenance and fault rectification

VEGASON 66 - 4 ... 20 mA/HART - four-wire