beautypg.com

VEGA VEGAPULS 65 (≥ 2.0.0 - ≤ 3.8) Foundation Fieldbus User Manual

Page 51

background image

Electromechanical data - version IP 66/IP 68 (1 bar)

Cable entry

-

Single chamber housing

1 x IP 68 cable gland M20 x 1.5; 1 x blind stopper
M

20 x 1.5

-

Double chamber housing

1 x IP 68 cable gland M20 x 1.5; 1 x blind stopper
M

20 x 1.5; 1 x blind stopper M16 x 1.5

Connection cable

-

Wire cross-section

0.5 mm² (AWG 20)

-

Wire resistance

< 0

.036 Ω/m

-

Tensile strength

< 1200 N

(270 lbf)

-

Standard length

5 m (16.4 ft)

-

Max. length

1000 m (3280 ft)

-

Min. bending radius

25 mm (0.984 in) with 25 °C (77 °F)

-

Diameter approx.

8 mm (0.315 in)

-

Colour - standard PE

Black

-

Colour - standard PUR

Blue

-

Colour - Ex-version

Blue

Indicating and adjustment module

Voltage supply and data transmission

through the sensor

Indication

LC display in dot matrix

Adjustment elements

4 keys

Protection

-

unassembled

IP

20

-

mounted into the sensor without cover

IP

40

Materials

-

Housing

ABS

-

Inspection window

Polyester foil

Power supply

Operating voltage

-

Non-Ex instrument

9 … 32 V DC

-

EEx-ia instrument

9 … 24 V DC

-

EEx-d instrument

16 … 32 V DC

Operating voltage with lighted indicating and adjustment module

-

Non-Ex instrument

12 … 32 V DC

-

EEx-ia instrument

12 … 24 V DC

-

EEx-d instrument

20 … 32 V DC

VEGAPULS

65 • Foundation Fieldbus

51

10 Supplement

28452

-EN

-090305