beautypg.com

VEGA VEGAPULS 62 (≥ 2.0.0 - ≤ 3.8) standpipe ver. 4 … 20 mA_HART four-wire User Manual

Page 27

background image

E

nter the requested parameter via the appropriate keys, save your

settings and jump to the next menu item with the [->] key.

T

he menu item "Extended setting" offers the possibility to optimise

VEGAPULS

62

for applications in which the level changes very

quickly. For this reason, select the function "Quick level change > 1 m/
min.
".

E

xtended setting

quick level change > 1 m/min.

Note:
S

ince with the function "Quick level change > 1 m/min." the generation

of an average value of the signal processing is considerably reduced,
false reflections by agitators or vessel installations can cause
measured value fluctuations. A false echo memory is thus recom-
mended.

H

igh sockets or vessel installations, such as e. g. struts or agitators as

well as buildup and weld joints on the vessel walls cause interfering
reflections which can impair the measurement. A false echo storage
detects and marks these false echoes, so that they are no longer taken
into account for the level measurement. A false echo memory should
be created with low level so that all potential interfering reflections will
be detected.

G

ating out of false signals

C

hange now?

P

roceed as follows:

1

M

ove from the measured value display to the main menu by

pushing [OK].

2

S

elect the menu item "Service" with [->] and confirm with [OK].

N

ow the menu item "False signal suppression" is displayed.

3

C

onfirm "False signal suppression - Change now" with [OK] and

select in the below menu "Create new". Enter the actual distance
from the sensor to the product surface. All false signals in this area
are detected by the sensor and saved after confirming with [OK].

Note:
C

heck the distance to the product surface, because if an incorrect (too

large) value is entered, the existing level will be saved as false signal.
T

he filling level would then no longer be detectable in this area.

Extended setting/Quick
level change

Gating out of false sig-
nals

VEGAPULS

62

standpipe version • 4 … 20 mA/HART four-wire

27

6 S

et up with the indicating and adjustment module PLICSCOM

28738

-EN

-090305