3 function chart – VEGA VEGASWING 61 - transistor (NPN-PNP) User Manual
Page 18
18
6 Setup
VEGASWING 61 • - transistor (NPN/PNP)
29225-EN-130409
"Function chart" (A - max. detection or overflow protection, B - min.
detection or dry run protection).
With this DIL switch (3) you can set the switching point to liquids
having a density between 0.5 and 0.7 g/cm³ (0.018 and 0.025 lbs/
in³). With the basic setting, liquids with a density of > 0.7 g/cm³ (0.025
lbs/in³) can be detected. In liquids with lower density, you must set
the switch to > 0.5 g/cm³ (0.018 lbs/in³). The specifications for the
position of the switching point relate to water - density value 1 g/cm³
(0.036 lbs/in³). In products with a different density, the switching point
will shift in the direction of the housing or tuning fork end depending
on the density and type of installation.
Note:
Keep in mind that foams with a density > 0.45 g/cm³ (0.016 lbs/in³)
are detected by the sensor. This can cause faulty switchings particu-
lary when used as dry run protection system.
6.3 Function chart
The following chart provides an overview of the switching conditions
depending on the set mode and the level.
Level
Switching status
Control lamp
Mode A
Overflow protection
closed
Green
Mode A
Overflow protection
open
Red
Mode B
Dry run protection
closed
Green
Mode B
Dry run protection
open
Red
Failure of the supply
voltage
(mode A/B)
any
open
off
Sensitivity adjustment (3)