3 instrument repair – VEGA VEGAVIB 63 Lock fitting ARV-VB63.2 - pressure range: -1…16 bar User Manual

Page 10

background image

10 Screw the pressure screw (11) with a view turns from top

into the mounting boss (7)

11 Clean the connection tube of the sensor and the surfaces

of the clamp (1) as well as the pressure screw (11)

carefully and remove grease, oil and dirt. Insert the sensor

into the lock fitting. Slide the tube into the requested

position and hold it

12 Make sure that the sensor is in the correct position (height.

The height adjustment of the sensor determines also the

switching point

13 Tighten the pressure screw (11) with a torque 128 ±10 Nm

(94 ±7 lbf ft)

14 Continue to turn the pressure screw (11) clockwise until the

hexagon surfaces of the pressure screw (11) and the

mounting boss (7) correspond (max. 1/6 turn)

15 Tighten the terminal screws (3) alternately. Make sure that

the gap between clamp and pressure screw is in parallel.

Tighten the terminal screws up to a torque of 4 ±1 Nm

(3 ±0.7 lbf ft).
Hence the clamp (1) is pressed against the tube and fixes

the tube of the sensor in this position

16 Fasten the locking angle (4) with the hold screw (6) and the

spring ring (5) laterally on the pressure screw (11)
Hence the pressure screw and the terminal screws are

protected against unintentional loosening

5.3 Instrument repair

If a repair of ARV-VB63.2 should be necessary, please send

the instrument to the following address:
VEGA Grieshaber KG; Abteilung Reparatur; Am Hohenstein

113; 77761 Schiltach / Germany

10

Lock fitting ARV-VB63.2 - – pressure range -1 … 16 bar

Maintenance and fault rectification

29754

-EN
-050804