.5 Mounting procedure, four-wire instruments

The electronics module is located in the electronics compart-
ment. The connection to the connection compartment is
provided via internal connection cables. The illustration shows
the position of the electronics compartment in the double-
chamber housing.

1

Fig. 4: Double chamber housing
1

Position of the electronics compartment

Proceed as follows:

1

Switch off power supply

2

Unscrew housing cover of the electronics compartment

3

Disconnect internal connection cables, on VEGASON 64,
65 and 66 also loosen the plug connector

4

Loosen the two holding screws with a screwdriver (Torx
size T 10 or slot size 4)

2

1

Fig. 5: Loosen the screws
1

Electronics module

2

Screws (2 pcs.)

5

Pull the previous electronics out by holding the opening
levers.