beautypg.com

6 sil conformity, 7 safety information for ex areas, 8 manufacturer declaration – VEGA VEGAVIB 67 - transistor output User Manual

Page 6

background image

2.6 SIL conformity

VEGAVIB 67 fulfills the requirements of functional safety

according to IEC 61508/IEC 61511. You can find further

information in the supplementary instructions manual "Safety

Manual - Functional safety (SIL) VEGAVIB series 60".

2.7 Safety information for Ex areas

Please note the Ex-specific safety information for installation

and operation in Ex areas. These safety instructions are part of

the operating instructions manual and come with the Ex-

approved instruments.

2.8 Manufacturer declaration

In conformity with DIN EN 60079-14/2004, para. 5.2.3, point

c1, the VEGAVIB 67 level sensor is suitable for use in zone 2.
The operator must use the instrument as it was intended to be

used and follow the specifications of the following documents:

l

this operating instructions manual

l

this manufacturer declaration (24647)

l

the corresponding installation regulations

Max. increase of the surface temperature during operation:

40 K (individual components in the instrument)
With an ambient temperature of 70 °C (158 °F) on the housing

and a process temperature of 70 °C (158 °F), the max. ambient

temperature during operation is 110 °C (230 °F).
Measures for maintaining explosion protection during oper-

ation:

l

Operate the instrument in the range of the specified

electrical limit values. Permissible supply voltage: see

"Technical data"

l

Mount and operate the instrument in such a way that no

ignition danger is expected by electrostatic charges. The

process fitting, the plastic-coated/covered probe part or the

housing are made of electrically non-conductive plastic

(depending on the version).

l

Make sure that the seal is mounted correctly between the

lower part of the housing and the cover. Screw the cover

on tightly.

l

Make sure there is no explosive atmosphere present if you

want to operate the instrument with opened cover

6

VEGAVIB 67 - with transistor output

For your safety

29297

-EN
-060508