Nominal range
Overload, max. pres-
sure6)
Overload, min. pressure
-0.2 … 0.2
bar/-20 … 20 kPa
30
bar/3000 kPa
-0.8
bar/-80 kPa
-0.5 … 0.5
bar/-50 … 50 kPa
35
bar/3500 kPa
-1
bar/-100 kPa
A
bsolute pressure
0 … 0.1
bar/0 … 10 kPa
15
bar/1500 kPa
0 … 1
bar/0 … 100 kPa
35
bar/3500 kPa
0 … 2.5
bar/0 … 250 kPa
50
bar/5000 kPa
0 … 5
bar/0 … 500 kPa
65
bar/6500 kPa
0 … 10
bar/0 … 1000 kPa
90
bar/9000 kPa
0 … 25
bar/0 … 2500 kPa
130
bar/13000 kPa
0 … 60
bar/0 … 6000 kPa
200
bar/20000 kPa
Reference conditions and influencing variables (similar to DIN EN 60770-1)
R
eference conditions according to DIN EN 61298-1
-
T
emperature
+
15 … +25 °C (+59 … +77 °F)
-
R
elative humidity
45 … 75 %
-
A
ir pressure
860 … 1060
mbar/86 … 106 kPa
(12.5 … 15.4
psi)
D
etermination of characteristics
L
imit point adjustment according to
IEC
61298-2
C
haracteristics
linear
R
eference installation position
upright, diaphragm points downward
I
nfluence of the installation position
<
0.2
mbar/20 Pa (0.003 psi)
Deviation determined according to the limit point method according to IEC 60770
7
)
A
pplies to digital HART interface as well as to analogue current output 4 … 20 mA.
S
pecifications refer to the set span. Turn down (TD) = nominal measuring range/set span.
D
eviation
-
T
urn down 1:1 up to 5:1
<
0.075 %
-
T
urn down up to 10:1
<
0.015 %
x TD
D
eviation with absolutely flush process fittings EV, FT
-
T
urn down 1:1 up to 5:1
<
0.05 %
-
T
urn down up to 10:1
<
0.01 %
x TD
7)
I
ncl. non-linearity, hysteresis and non-repeatability.
S
upplement
VEGABAR
74 - 4 … 20
mA/HART
29
28432
-
EN