5 diagnostics, 1 maintenance, 2 failure rectification – VEGA VEGABAR 40 4 … 20 mA User Manual

Page 23: Fault signals, Troubleshooting

background image

VEGABAR 40 (4 … 20 mA)

23

Diagnostics

5 Diagnostics

5.1 Maintenance

VEGABAR process pressure transmitters are
maintenance-free.

5.2 Failure rectification

Fault signals

VEGABAR provides maximum reliability with
its self-test and continuous self-monitoring.
VEGABAR diagnostics distinguishes between
atypical process conditions and faults in the
VEGABAR.

Atypical process conditions

Exceeding or falling short of measuring
range limits (fault signal extinguishes when
the measured value is again in the measuring
range).

Failure in VEGABAR

Failure in the electronics, interference or
damage of the measuring cell.

The following table assists in analysis of fault
signals.

Reason

Fault signal via

DOT-matrix

Bar graph

Current

Digital indication

output

Clearly

"OPERATE

Bar graph 0 %

outside

????

or 100 %

Current

measuring

bar“

digital value

value

range

flashes

3.6 mA
or
22 mA

Overload

Bar graph 0 %

range

or 100 %,

of the

digital value, four

meas. cell

flashing seg-
ments "- - - -“

Failure in

all segments

VEGABAR

flash

On instruments with menu-driven adjustment
with additional functions, the possible rea-
sons are displayed under the menu
item"Diagnostics no.“.

Diagnostics no.

Meaning

1

Connection to CID converter inter-
rupted

2

Frequency signal of the capacitor not
within the limit values

3

Frequency signal of the reference
capacitor not within the limit values

4

Frequency signal temperature not
within the limit values

7

Communication error to EEPROM

9

Error in EEPROM CRC-checksum

11

Process fitting or electronics unit has
been exchanged (appears for approx.
20 s after switching on for the first
time after the exchange)

Troubleshooting

If the displayed value does not correspond
to the level in the vessel or to the process
pressure, the following measures must be
taken:
- check the pressure compensation (only

with gauge pressure measuring ranges)

- check the electrical connection.