3 technical data – VEGA Capacitive electrodes EK … with signal conditioning instrument User Manual
Page 8
8
Capacitive electrodes EK with signal conditioning instrument
Product description
1.3 Technical data
Housing
Housing material
plastic PBT (Polyester) or Aluminium
plastic coated
Protection
- plastic housing
IP 66
- Aluminium housing
IP 66 and 67 (meets both protections)
Cable entry
1 pce. M20 x 1,5
Terminals
for max. 1,5 mm
2
cross-section area of
conductor
Mechanical connection
Material
1.4435 (316 L)
Thread
G
3
/
4
A or
3
/
4
“ NPT
G 1 A or 1“ NPT
Flange
flange versions, plated
Electrode
Material
EK 11
1.4435 (316 L)
EK 21
steel (St 37), 1.4435 (316 L)
EK 31
1.4401 (316 L)
EK 24, 42 1.4571 (316 L)
Length
- rod
max. 3 m
- cable
max. 20 m
Isolation
see "Isolating materials"
Max. tensile strength
- EK 31
3 KN
- EK 42
3 KN
Ambient conditions
Ambient temperature on the housing
-40°C … +80°C
Medium temperature
see "Product temperature and operating
pressure"
Storage and transport temperature
-40°C … +80°C
Operating pressure
see "Product temperature and operating
pressure"
Oscillators E14, E15, E17, E18
Protection class
II
Overvoltage category
III
Meas. frequency
see table on the following page
Capacitance ranges
see table on the following page
Supply voltage
12 … 36 V DC (powered by signal conditioning
instrument)
Potential separation
min. 500 V DC (except E14)