6 safety-related characteristics, 6safety-related characteristics, Supplement – VEGA VEGAPULS 41 4 … 20 mA; HART compact sensor User Manual
Page 64: With t, 1 year, 5 years pfh [1/h, Failure reaction time t, 60 sec
64
VEGAPULS 41 – 4 … 20 mA
26620-EN-041227
0
1
2
3
4
5
years
T
proof
= 1 year
T
proof
= 3 years
T
proof
= 5 years
0,00·10
-2
0,20·10
-2
0,40·10
-2
PFD
avg
0,30·10
-2
0,50·10
-2
0,10·10
-2
Supplement
6
Safety-related characteristics
The failure rate of the electronics and the antenna system was determined by an FMEDA (Fail-
ure Mode, Effects and Diagnostics Analysis) acc. to IEC 61508. The number values are based
on an output current tolerance of max. 2 %.
Architecture 1oo1D
SIL 2
VEGAPULS 41
Overfill protection
Dry run protection
Individual level
VEGAPULS 42
VEGAPULS 43
VEGAPULS 44
VEGAPULS 45
HFT
0
SFF
> 88%
PFD
avg
1)
< 0.10 •10
-2
with T
Proof
= 1 year
2)
< 0.50 •10
-2
with T
Proof
= 5 years
PFH [1/h]
3)
< 0.22 •10
-6
Failure reaction time
T
Reaction
= 60 sec
Max
Min
1)
PFD
avg
this value correlates almost linearly to the operating time. It is only valid for the corresponding
selection circuit.
2)
T
Proof
is the interval after which a periodically recurring complete function test (to check the safety function)
must be carried out.
3)
PFH is valid for the stated Failure reaction time T
Reaction
of the measuring system. This means that the failure
tolerance time of the complete system must be higher than T
Reaction
.