beautypg.com

6 safety-related characteristics, 6safety-related characteristics, Supplement – VEGA VEGAPULS 41 4 … 20 mA; HART compact sensor User Manual

Page 64: With t, 1 year, 5 years pfh [1/h, Failure reaction time t, 60 sec

background image

64

VEGAPULS 41 – 4 … 20 mA

26620-EN-041227

0

1

2

3

4

5

years

T

proof

= 1 year

T

proof

= 3 years

T

proof

= 5 years

0,00·10

-2

0,20·10

-2

0,40·10

-2

PFD

avg

0,30·10

-2

0,50·10

-2

0,10·10

-2

Supplement

6

Safety-related characteristics

The failure rate of the electronics and the antenna system was determined by an FMEDA (Fail-
ure Mode, Effects and Diagnostics Analysis) acc. to IEC 61508. The number values are based
on an output current tolerance of max. 2 %.

Architecture 1oo1D

SIL 2

VEGAPULS 41

Overfill protection

Dry run protection

Individual level

VEGAPULS 42
VEGAPULS 43
VEGAPULS 44
VEGAPULS 45

HFT

0

SFF

> 88%

PFD

avg

1)

< 0.10 •10

-2

with T

Proof

= 1 year

2)

< 0.50 •10

-2

with T

Proof

= 5 years

PFH [1/h]

3)

< 0.22 •10

-6

Failure reaction time

T

Reaction

= 60 sec

Max

Min

1)

PFD

avg

this value correlates almost linearly to the operating time. It is only valid for the corresponding

selection circuit.

2)

T

Proof

is the interval after which a periodically recurring complete function test (to check the safety function)

must be carried out.

3)

PFH is valid for the stated Failure reaction time T

Reaction

of the measuring system. This means that the failure

tolerance time of the complete system must be higher than T

Reaction

.