6 instrument modification, 1 exchange of adjustment modules – VEGA VEGABAR 44 4 … 20 mA User Manual
Page 25
VEGABAR 44 (4 … 20 mA)
25
Instrument modification
6 Instrument modification
6.1 Exchange of adjustment
modules
The modular construction of VEGABAR allows
retrofitting, interchange or removal of the
adjustment and display modules. Data previ-
ously saved (e.g. adjustment values) are not
stored in the adjustment module but in an
EEPROM of the sensor electronics and
therefore readjustment is not necessary. The
connection of the modules is made via a 4-
pole plug connection.
Exchange of the adjustment module
Removal of the adjustment module
•
Separate VEGABAR from the power
supply
•
Loosen the screws on the upper side of
the housing and remove the cover or the
display module
•
Loosen connection cables from the termi-
nals, if necessary pull out the plug con-
nection of the display module
•
Remove the two screws of the adjustment
module
•
Remove adjustment module and pull out
plug connection
Insertion of an adjustment module
•
Plug the plug connection of the new ad-
justment module into the plug-in socket of
the sensor electronics
•
Fasten the new adjustment module
•
Reconnect the connection cables, if
necessary, also the cable from the
display module
•
Close cover or display module of VEGA-
BAR
•
Reconnect VEGABAR to power supply
VEGABAR
Module
without
adjustment
Module for
menu-driven
adjustment
Module for
basic
functions
Display
module