5
Product description
1.6 Technical data
Power supply
Operating voltage/Power consumption
U
nom
= 20 … 250 V AC, 50/60 Hz/approx. 12 VA
= 20 … 72 V DC
Output or measured data input
Number
1 x 24 V output or digital measured data input
for up to 15 sensors
Type of measured data input
active two-wire input, digital (HART
®
protocol),
sensors are powered by VEGASTAB 692 MD
Sensor type
up to 15 VEGA two-wire loop-powered HART
®
sensors in Multidrop mode
(pressure transmitters, ultrasonic and radar
sensors, see chapter "1.5 Sensor choice")
Output voltage
approx. 17 … 24 V DC
Output constant current
max. 60 mA (15 sensors with 4 mA each)
Current limitation
at approx. 200 mA, short-circuit proof
Connection cable
2-wire standard cable (screening is
recommended)
max. cable length
1000 m
max. resistance per wire
15 Ohm
Serial interface (RS 232)
Number
1 x RS 232 via VEGACONNECT 3
Function
communication via modem or PC for data
recording with Visual VEGA (VV) or
sensor adjustment with VEGA Visual Operating
(VVO)
max. cable length
15 m
Interface parameters
9600 baud, no parity, 1 stop bit
Protocol
VEGA specific
Indicating elements
LED in front plate
green: power supply and output voltage on
Ambient conditions
Permissible ambient temperature
-20°C … +60°C
Storage and transport temperature
-40°C … +80°C
Electrical connection
Screw terminals
max. 1.5 mm
2
RS 232 interface (VEGACONNECT 3)
Sub-D 9-pole
Electrical protective measures
Instrument
IP 30
Plug-in socket
IP 20
Protection class
II
Overvoltage category
II