6.1 Cleaning

The housing may be cleaned with a cloth dampened with a mild detergent if
necessary. Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean
the housing or control panels.


6.2 Troubleshooting

The following table lists common problems and possible causes and remedies. If
the problem persists, contact Ohaus or your authorized dealer.


TABLE 6-1. TROUBLESHOOTING.

Symptom Possible

Cause(s)

Remedy

Scale will not turn on

No power to scale
Battery power used up

Verify connections and voltage.
Connect power and charge the battery

Poor accuracy

Improper calibration • Unstable
environment • Internal pressure
build-up

Perform calibration • Move scale to
suitable location • Relieve internal
pressure by loosening rear leveling feet

Unable to calibrate

LFT Menu set to ON
Unstable environment
Incorrect calibration mass

Set LFT menu to OFF (see section 5.1)
Move the scale to suitable location
Use correct calibration mass

Cannot display weight in
desired weighing unit

Weighing unit not enabled

Enter menu and enable unit

Cannot change menu
settings

LFT Menu set to ON

Set LFT menu to OFF (see section 5.1)

Scale displays Lo bAt

Battery is discharged

Connect power and charge the battery

Battery fails to charge
fully

Battery is defective

Have battery replaced by authorized
service dealer

Err 1

Weight on pan exceeds capacity

Remove weight from the pan

Err 2

Underload limit exceeded • Pan
was removed prior to power on

Install pan and re-zero

Err 9, Err 21

Factory data corrupted

Contact Ohaus or an authorized service
agent