Calibration

For best results, calibrate the scale at regular intervals. This is espe-

cially important if the scale is in use for prolonged periods.

1. Allow the scale to warm up and stabilize for two

minutes before commencing the calibration

procedure.

2. Press and hold the “Units/Cal” key until “CAL”

flashes on the display.

3. Press the “On/Off/Zero” key to capture the scale

zero. The platform should be empty and the

display shows “-C-”

4. After the zero is captured, the display will show

“xxx”. The value “xxx” is the span calibration

weight.

5. Place the calibration weight on the platform and

press “On/Off/Zero” key. The display will again

show “-C-”.

6. After span capture, the display will return to the

normal weighing mode.

NOTE: If “CAL E” appears on the display, a calibration

error has occurred and the process should be re-

peated. The message “CAL E” will appear if the

calibration steps are not followed in the correct order

or if an incorrect calibration weight was applied.