beautypg.com

Ohaus BW WASHDOWN SCALES Manual multi User Manual

Page 11

background image

BW SERIES COMPACT BENCH SCALE EN-

6. “

g” s dsplayed. Ths s the Calbraton Weght Unt. Press -Yes to

accept (or press

-No to swtch to “lb”, then press -Yes) and move on to the

next parameter. Refer to the followng chart for the next steps.

Selected calibration weight unit: g

Scale Capacty

1.5 kg / 3 lb

3 kg / 6 lb

6 kg / 15 lb

15 kg / 30 lb

Dsplayed Settng

1500

3000

6000

15000

2/3 of Capacty

1000 / 1 kg

2000 / 2 kg

4000 / 4 kg

10000 / 10 kg

Full Capacty

1500 g

3000 g

6000 g

15000 g

Selected Calibration weight unit: lb

Scale Capacty

3 lb / 1.5 kg

6 lb / 3 kg

15 lb / 6 kg

30 lb / 15 kg

Dsplayed Settng

3000

6000

15000

30000

2/3 of Capacty

2000 / 2 lb

4000 / 4 lb

10000 / 10 lb

20000 / 20 lb

Full Capacty

3000 / 3 lb

6000 / 6 lb

15000 / 15 lb

30000 / 30 lb


7. “

xxxx” s dsplayed. Ths s the Full Scale Capacty of the scale. Make

sure the dsplayed value matches the scale model capacty then press

-Yes

to accept and move on to the next parameter.

8. “

- - - - - -” s dsplayed. Make sure the pan s empty and stable at ths tme

then press

-Yes. The dsplay wll count down from 5 to 0 as the scale zero

s regstered.

9. “

xxxx” s dsplayed. Ths s the Calbraton Pont correspondng to 2/3 of

the Full Scale Capacty value. Place the ndcated calbraton weght on

the center of the pan. Wth the scale stable, press

-Yes. The dsplay wll

count down from 5 to 0 as the scale regsters the calbraton weght.

10.

“xxxx” s dsplayed. Ths s the Calbraton Pont correspondng to the Full

Scale Capacty value. Place the ndcated calbraton weght on the center

of the pan. Wth the scale stable, press

-Yes. The dsplay wll count down

from 5 to 0 as the scale regsters the calbraton weght.

11. “

SAVE” s dsplayed when calbraton was regstered. Remove the

calbraton weght, then press

-Yes. The scale wll dsplay a self-test then

return to normal weghng mode. Or, press

-No to abort calbraton wthout

savng any changes.

Note: At any tme durng the calbraton process, long-press ON/UNITS

OFF

to turn

off the scale and abort the process wthout savng any changes.