The temperature is 
above or below the 
proper range for the 
system. Move to an 
area within the 
temperature range 
indicated for test strip 
use in the test strip 
package insert, wait fi ve minutes, and 
repeat the test. Do not artifi cially heat or 
cool the meter.
The batteries are 
almost out of power. 
Change the batteries 
now.
Blood or control solution 
was applied to the test 
strip before the fl ashing 
drop symbol appeared 
on the display. Discard 
the test strip and repeat 
the test.
An electronic error 
occurred or, in rare 
cases, a used test strip 
was removed and 
reinserted. Turn the 
meter off and on, or 
take the batteries out 
for 20 seconds and 
reinsert them. Perform a blood glucose 
or control test. If the problem persists, 
call Roche.