beautypg.com

Display messages and troubleshooting (continued) – Accu-Chek Compact Plus User Manual

Page 80

background image

Chapter 5: Maintenance and Troubleshooting

78

ACCU-CHEK Customer Care Service Center 1-800-858-8072

Display Messages and Troubleshooting (continued)

The bar code information stored in

the meter is lost. Slide and release

to turn off the meter. Open and

close the drum door. Make sure a test

drum is loaded in the meter.

There is no test drum in the meter, or

the bar code on the drum is damaged

or missing. Make sure your test drum

has two bar codes on it. Insert a new

test drum.

The test reminder is turned on. The

test reminder symbol flashes on the

display and the meter beeps, if the

beeper is turned on.