Humidifier performance – ResMed S9 H5i User Manual
Page 10
English
6
Notes:
•
The manufacturer reserves the right to change these specifications without notice.
•
The temperature and relative humidity settings displayed for ClimateLine or ClimateLine
MAX
are not
measured values.
•
Check with your clinician/service provider before using the SlimLine air tubing with devices other
than the S9 or H5i.
Humidifier performance
The following settings have been tested at 22°C ambient temperature:
Cleanable water tub
Injection molded plastic, stainless steel and silicone seal
Standard water tub
Injection molded plastic, aluminium and thermoplastic
elastomer
Operating temperature
+5ºC to +35ºC
Operating humidity
10–95% non-condensing
Storage and transport temperature
-20ºC to +60ºC
Storage and transport humidity
10–95% non-condensing
ClimateLine air tubing
Flexible plastic and electrical components, 2 m, 15 mm
inner diameter
ClimateLine
MAX
air tubing
Flexible plastic and electrical components,1.9 m, 19 mm
inner diameter
Heated tubing temperature cut-out
41°C
IEC 60601-1 classification
Class II (double insulation), Type BF
Electromagnetic compatibility
Please refer to the S9 Information Guide for details
CPAP mask
pressure, cm H
2
O
RH output %
Nominal system output AH
a
, BTPS
b
a.
AH - Absolute Humidity in mg/L.
b. BTPS - Body Temperature Pressure Saturated.
Setting 3
Setting 6
Setting 3
Setting 6
4
90
100
10
18
10
95
100
11.5
21
20
95
100
11
18
25
100
100
12
13.5