Technical specifications – ResMed S9 H5i User Manual
Page 9

English
5
Technical specifications
Displays error message:
High leak, please check system setup and all connections
There is excessive leak.
Adjust position of mask and headgear.
Air tubing is not connected properly. Connect firmly at both
ends.
The H5i flip lid is not latched correctly.
Close the flip lid ensuring that it clicks into place.
Displays error message:
Please close H5i flip lid, attach tube and press any key
H5i flip lid is not closed.
Close the flip lid ensuring that it clicks into place.
Air tubing is not connected properly.
Connect firmly at both ends.
Water splashing on your face from the H5i
The water tub is overfilled.
Check that the water level is below the maximum water
level mark.
Condensation is forming in the air tube
and mask.
Turn the H5i or relative humidity setting down via the
menu.
Leaking water tub
The water tub may be damaged or
cracked.
Contact your service provider for a replacement.
The cleanable water tub is not
assembled correctly.
Check for damage and reassemble the cleanable water tub
correctly.
Air feels too warm/cold in the mask
The temperature of the ClimateLine or
ClimateLine
MAX
air tubing is set too
high/low.
Turn up/down the heated tubing temperature via the menu.
Maximum heater plate temperature
65°C
Temperature cut-out
74°C
Maximum gas temperature
41°C
Nominal dimensions (L x W x H)
Docking station and water tub: 153 mm x 145 mm x 86 mm
Weight (Standard water tub)
Docking station and unfilled water tub 0.69 kg
Weight (Cleanable water tub)
Docking station and unfilled water tub 0.79 kg
Water capacity
To maximum fill line 380 mL
Docking station
Flame retardant engineering thermoplastic, aluminium
Problem/Possible cause
Solution