LF2020/LF2090 Sit-To-Stand Patient Lift Owner’s Manual

13

Proper care and maintenance is essential to

keeping your Lumex patient lift in safe operating

condition.

In addition to inspecting the unit before each use,

periodic maintenance checks should be done as

needed. The lift should be thoroughly inspected by

a person qualified to recognize any signs of wear

and tear, and looseness of bolts or parts. Replace

any worn parts immediately.

We further recommend that this unit be serviced at

least once annually by an authorized dealer.

Warning: It is extremely important that the
patient lift be inspected before each use.
Ensure that all hardware and accessories
are secure and that the actuator is
functioning properly. Failure to do so could
result in patient / attendant injury or
damage to your patient lift. Torn, cut, frayed
or broken slings could fail, resulting in
serious injury. Use only slings in good
condition. Inspect before each use. Destroy
and discard old, worn and unusable slings.
When you believe a component or part is
not functioning properly, immediately
contact your authorized Graham-Field
Health Products dealer, as a potentially
hazardous condition could exist.

Before each use

Before each use of the patient lift, ensure that:

• All nuts and bolts are tight

• The lift moves freely (with caster brakes off )

• Rear caster brakes operate properly

• Base adjustment mechanism operates easily

Lubrication

Under normal use, your Lumex patient lift was

designed with minimum maintenance in mind;

however, a six (6) month check and lubrication

should ensure that your patient lift will remain in

peak operating condition. The casters MUST roll

and swivel freely. Lightweight waterproof grease

(such as automobile lubricant) should be applied

to the ball bearings of the casters every six (6)

months. Apply grease more frequently if the

casters are exposed to periods of excess

moisture. Lubricate all pivot points with

lightweight oil. Clean patient lift of all excess oil

and grease.

Before use of the patient lift, and every month

thereafter, place a drop of oil on the following areas:

• The point where mast and boom connect

• Caster axles

• The point where actuator and boom connect

Cleaning

Regular cleaning is recommended. A soft cloth,

dampened with water and a mild NON-ABRASIVE

detergent (household cleaner, soapy water, or

hospital grade disinfectant) is all that is needed

to clean your patient lift. Automobile wax or

furniture polish will help maintain the finish over

a long period of time.

DO NOT wash under water pressure or steam clean.

Checking for damage and wear

It is extremely important to frequently check the

areas of the patient lift that are under a great

deal of stress for damage and wear. These areas

include slings, spreader bars, spreader bar

suspension and mounting hardware, pump

mounting hardware and all pivot points. Inspect

for signs of cracking, fraying, deformation, and

any loose components. Replace any defective

parts immediately, and make certain that the

patient lift is not used until repairs are made.

Contact your authorized Graham-Field Health

Products dealer for replacement parts.

Care of sling

Lumex slings are easy to maintain. They are

washable at 150° F. The slings can be dried by

using a dryer on LOW heat or by hanging.

Note: Slings should not be bleached!

CARE AND MAINTENANCE