LF1050 Patient Lift Owner’s Manual

15

TROUBLESHOOTING

Symptom

Possible Cause

Action

Control box is not responding and
relay is not clicking

Battery discharged
Battery charger defective
Battery defective
Hand control defective

Charge battery
Replace battery charger
Replace battery
Replace hand control

Actuator does not run,
but relay is clicking

Actuator plug not properly

seated

Actuator defective
Control box defective

Push actuator plug firmly into control

box

Replace actuator
Replace control box

Actuator goes up or down, but not
both

Hand control defective
Control box defective

Replace hand control
Replace control box

Cannot lift patient up and battery

indicator shows no GREEN light

Battery discharged

Charge the batteries

Lift does not work and the battery

indicator shows full charge

Check the connection between the

actuator and the control box

Check connection between Pendant &

control box

Lift does not work and the battery

indicator shows no light

Make sure the emergency stop button

is released

Check the connection of the batteries

in control box

“Stops-and-Runs” while lifting or

lowering patient

Check socket for pendant plug-in
Check pendant cable

Battery Indicator shows no light

when charger is connected to
power supply

Check the emergency stop button is

released.

Check charger plug for damage.
Check charger cable for damage.

Lift does not work. Battery indicator

shows full charge. Notice “click”
sound from control box when UP
or DOWN button on hand control
was pressed.

Actuator problem.

Warning: If you experience a problem with your lift and
are unable to service it yourself, contact your Lumex dealer.