LF1050 Patient Lift Owner’s Manual

12

CHECKING THE BATTERY

1. Check batteries by pressing the Battery Testing

Button (blue circle with a battery sign) on the
control box.

2. Batteries are fully charged when all lights on the

Battery Indicator are "ON".

3. Charge batteries when Indicator shows only one

“GREEN” light.

4. Do not use the lift when no “GREEN” light is

shown. Charge the batteries immediately.

5. Replace batteries when frequent charging is

needed. (Reorder #: DPL650BAT)

CHARGING THE BATTERIES

1. Ensure the power is switched "ON" (the red

“RESET” button should be up).

2. Insert charging plug into charging port on the

control box.

3. Plug charger to power supply.
4. All lights of battery indicator should be "ON"

while charging.

5. It takes approximately 2-3 hours to fully charge

the batteries from one green light.

6. It takes approximately 7-9 hours to fully charge

the batteries from the red light.

7. Do not leave the batteries on charging for an

extended period of time. This will shorten the
life of the batteries.

8. Do not let the batteries run down to the last red

light. This might shorten the life of the batteries
or damage the batteries.

9. Unplug the charger first before using the lift.

THE LF1050 PATIENT LIFT WILL NOT OPERATE
WHILE CHARGING.

IMPORTANT NOTES ON CHARGING

1. Push lift to charging location and charge the

batteries with the charger provided. Avoid
unplugging hand control and motor from control
box. Frequent plug/unplug of the hand control
and motor into/from the control box may
damage the control box.

2. Do not charge batteries over 12 hours.

Red

Amber

Green

Green

Green

ON

ON

ON

ON

ON

FULLY CHARGED

ON

ON

ON

OFF

OFF

NEEDS CHARGE

ON

ON/OFF

OFF

OFF

OFF

CHARGE IMMEDIATELY. DON’T USE THE LIFT

OFF

OFF

OFF

OFF

OFF

BAD BATTERIES OR BAD CONNECTION

Know What The Battery Indicator Lights Are Telling You

Warning Buzzers
The control box has a buzzer that provides you with information on the status of the batteries and the lift.
To stop the buzzer, push down the RESET button or press the Battery Indicator.

BUZZER TYPE

INDICATION

NOTE

2 beeps/sec

Low Battery (red / amber lights on)

Charge batteries immediately

3 beeps/sec

Low Battery (red light on)

Charge batteries immediately

6 beeps/sec

Very Low Battery (no light on)

Charge batteries immediately

3 beeps every
other second

Bad Battery (no light on)

Replace batteries

3 long beeps

Overloaded. Batteries are charged
but lift does not lift.

Exceeded maximum load capacity