8. Move the holder (sensor) center so that the flat portion center at the top of the D.SENS (CH1) input waveform and the ‘’H” center of
HHOUT (CH2) coincide.
9. Tighten the fixed screw to fix the disc table, then press the
[CLEAR]
button.
10. The disc table rotates in the clockwise direction. Measure the waveform and make sure that the flat portion center at the top of the
D.SENS (CH1) input waveform and the ‘’H” center of HHOUT (CH2) conform.
Note:
During the adjustment mode, the rotational direction is switched each time the
[CLEAR]
button is pressed.
Pressing the
[CHECK]
button enters the loading mode which will be described later. Pressing the
[CLEAR]
button rotates the disc table again.
Contets of the fluorescent indicator tube's display in the loading mode
HHOUT (CH2) waveform
D.SENS (CH1) waveform
Flat portion center of
D. SENS (CH1) waveform
Fixed screw
Holder (sensor)
51
O N
I N
O F F
ALL DISCS
DISC
CD1
Loading OUT
SW condition
Loading
direction
Loading IN
SW condition