Labs cryostorage specifications – Taylor TW-357 User Manual
Page 7

TW-357 REV E
7
LABS Cryostorage Specifications
Liquid nitrogen at atmospheric pressure weighs 1.78 lb./liter (0.8 kg/liter). All Taylor-Wharton Gas
Equipment cryostorage systems are designed to support the full weight of liquid nitrogen and a complete
stainless steel or aluminum inventory control system with boxes and specimens.
Dimensions
80K
40K
20K
External
Operating
Height in
42.5
42.5
42.5
(Top of Step to Lid opening)
mm
1080
1080
1080
Step
Height
in
11
11
11
mm
279
279
279
Overall
Height
in.
60
60
60
(Top
of
Control
Interface) mm
1524
1524
1524
Usable
Height,
Internal
in.
30.0
30.0
30.0
mm
762.0
762.0
762.0
Outside
Diameter in
59.5
45.0
34.0
mm
1511
1143
863.6
Internal
Working
Diameter
in
55.0
40.5
29.5
mm
1397
1029
750
Neck
Opening
in
24.5
18.0
13.0
mm
622
457
330
Capacity
LN2
Capacity
L
1350
606
407
Power Supply
0
VAC 16.5
16.5
16.5
Evaporation Rate
1
L/day
15.0
9.0
8
Weight, Empty
lb
1550
920
650
kg
703
417
295
Maximum Gross Weight
2
lb.
3956
2000
1375
kg
1794
907
624
Inventory Control System Specifications
No 5.0 x 5.0 Racks
3
58
30
14
No.
Shelves/Rack
13
13
13
No. 3.0 x 3.0 Racks
4
12
8
4
No.
Shelves/Rack
13
13
13
Vial Capacity, 2 ml
5
79300
41600
19500
Blood
Bag
(Fenwall
4R9951)
2100
1204
560
Blood
Bag
(Fenwall
4R9953)
1505
820
400
Blood
Bag
(Fenwall
4R9955)
1050
590
290
Footnotes:
0
This is the power supply for the standard battery backup version. The Kryos version (No Battery) uses a 24 VAC power supply.
For power outlets other than North American 110 VAC contact Taylor-Wharton Customer Service.
1
Evaporation rate is nominal. Actual rate may be affected by the nature of the contents, atmospheric
conditions, container history, and manufacturing tolerances.
2
Includes the empty weight and total LN2 capacity weight.
3
5.0 in. x 5.0 in. (127 mm x 127 mm) 100 cell box.
4
3.0 in. x 3.0 in. (76 mm x 76 mm) 25 cell box.
5