beautypg.com

Teledyne OXYGEN DETECTOR 319R User Manual

Page 13

background image

9

319R

Maintenance 3.0

TELEDYNE BROWN ENGINEERING

Analytical Instruments

Maintenance

Replacing the Fuse

If the unit will not turn on when plugged in and switched ON (via the

power switch), or you suspect a blown fuse, replace the fuse using the
following procedure:

1. Disconnect the AC power and place the power switch located

on the rear panel in the OFF position.

2. The fuse is located next to the power cord receptacle in the

upper left-hand corner of the rear panel. Remove the power
cord from the receptacle.

3. Insert a small flat-blade screwdriver into the slot in the

receptacle wall nearest the fuse and gently pry open the fuse
receptacle. The fuse holder will slide out. The fuse in use is
visible in the clip. To open the spare fuse compartment, push on
one end until it slides out.

4. Remove the bad fuse and replace it with a 5X20 mm 1A, 250V

fast acting fuse for AC units (TBE P/N F-1189 ), or 1/2A fast
acting fuse for DC units (P/N F-51).

Figure 5-1: Replacing the fuse (AC power).

This manual is related to the following products: