Teledyne OT-3 User Manual

Page 41

background image

Trace Oxygen Analyzer

Maintenance

Teledyne Analytical Instruments

33

CAUTION:

READ THE SECTION CELL WARRANTY
CONDITIONS
, BELOW, BEFORE REPLACING THE
CELL.

After replacing the Micro-fuel Cell, the analyzer must be

recalibrated. See Calibration in Chapter 4.

5.2.2 Ordering and Handling of Spare Sensors

To have a replacement cell available when it is needed, TAI

recommends that one spare cell be purchased when the current cell 's
warranty period is approximately two thirds over.

CAUTION:

DO NOT STOCKPILE CELLS. THE WARRANTY
PERIOD STARTS ON THE DAY OF SHIPMENT. FOR
BEST RESULTS, DO NOT ORDER A NEW SPARE
CELL TOO SOON.

The spare cell should be carefully stored in an area that is not

subject to large variations in ambient temperature (75 °F nominal), and
in such a way as to eliminate the possibility of incurring damage.

CAUTION:

DO NOT DISTURB THE INTEGRITY OF THE CELL
PACKAGE UNTIL THE CELL IS TO ACTUALLY BE
USED. IF THE CELL PACKAGE IS PUNCTURED AND
AIR IS PERMITTED TO ENTER, CELL-LIFE WILL BE
COMPROMISED.

WARNING:

THE SENSOR USED IN THE MODEL OT-3 CONTAINS
AN ELECTROLYTE WHICH INCLUDES SUBSTANCES
THAT ARE EXTREMELY HARMFUL IF TOUCHED,
SWALLOWED, OR INHALED. AVOID CONTACT WITH
ANY FLUID OR POWDER IN OR AROUND THE UNIT.
WHAT MAY APPEAR TO BE PLAIN WATER COULD
CONTAIN ONE OF THESE TOXIC SUBSTANCES. IN
CASE OF EYE CONTACT, IMMEDIATELY FLUSH
EYES WITH WATER FOR AT LEAST 15 MINUTES.
CALL A PHYSICIAN. (SEE APPENDIX, Material Safety
Data Sheet—MSDS).

5.2.3 Removing the Micro-fuel Cell

To remove a spent or damaged Micro-fuel Cell:

1. Disconnect the Power Source from the instrument.