beautypg.com

1 features – Teledyne MX300 User Manual

Page 11

background image

Portable Oxygen Monitor

Introduction

Teledyne

Analytical

Instruments

3

NOTE: IF THE LOW ALARM CONDITION IS TRIGGERED WITH

THE OXYGEN LEVEL STAYING ABOVE THE SAFETY LIMIT
OF 18%, THE ALARM BUZZER CAN BE DEACTIVATED BY
PRESSING THE ALARM SILENCE KEY. IF ALARM CONDI-
TION IS NOT CORRECTED WITHIN ABOUT 115 SECONDS,
THE ALARM BUZZER WILL REACTIVATE. PRESSING THE
ALARM SILENCE KEY AGAIN WILL SILENCE THE ALARM
BUZZER ONCE AGAIN FOR ABOUT 115 SECONDS, WITH
THE BUZZER REAACTIVATING IF THE ALARM CONDI-
TIONS PERSIST.

NOTE: IF THE LOW ALARM CONDITION IS TRIGGERED WITH

THE OXYGEN LEVEL FALLING BELOW 18%, THE

SENSOR SIGN WILL FLASH ALONG WITH THE GREEN

LIGHT, AND THE BUZZER WILL ACTIVATE. IN THIS CON-
DITION, THE ALARM BUZZER CANNOT BE SILENCED.

WHILE HOLDING DOWN THE ALARM SILENCE KEY MAY
MOMENTARILY SILENCE THE ALARM BUZZER, AUDIBLE
ALARM WILL RETURN UPON DISENGAGING THE ALARM
SILENCE KEY. HOLDING THE KEY FOR LONGER DURA-
TIONS WILL INITIATE THE “STUCK KEY” ALARM, WHICH
CAN BE DEACTIVATED ONLY VIA REMOVING THE BAT-
TERIES FROM THE ANALYZER. ANALYZER SHOULD BE
RECALIBRATED UP ON REPOWERING.

1.1 Features

The following features are standard on the MX300 Monitor:

Large easy to read 3½-digit LCD (see options)

Automatic LCD back lighting upon key press

Microprocessor controlled

Up/Down control keys on the front panel

Sensor fail/disconnect alarm indicator (audible and visual)

Two adjustable concentration alarms with user-defined set
points

Alarm Silence key

2000 operating hours from 3 AA alkaline batteries

Battery status indicator

Battery test function

Stand for upright tabletop deployment

This manual is related to the following products: