Trace Oxygen Analyzer

Installation 3

3-9

Teledyne Analytical Instruments

Note: The remote probe connections (paragraph 3.3.9) provide

signals to ensure that the zero and span gas valves will be
controlled synchronously. If you have the –C Internal valve
option—which includes additional zero and span gas inputs—
the 3020T automatically regulates the zero, span and sample
gas flow.

3.3.6 Range ID Relays

There are four dedicated RANGE ID CONTACT relays. The first

three ranges are assigned to relays in ascending order—Low range is as-
signed to RANGE 1 ID, Medium range is assigned to RANGE 2 ID, and
High range is assigned to RANGE 3 ID. RANGE 4 ID is reserved for the
Air Cal Range (25%).

3.3.7 Network I/O

A serial digital input/output for local network protocol. At this printing,

this port is not yet functional. It is to be used in future versions of the instru-
ment.

3.3.8 RS-232 Port

The digital signal output is a standard RS-232 serial communications

port used to connect the analyzer to a computer, terminal, or other digital
device. The pinouts are listed in Table 3-3.

Table 3-3: RS-232 Signals

RS-232 Sig

RS-232 Pin

Purpose

DCD

1

Data Carrier Detect

RD

2

Received Data

TD

3

Transmitted Data

DTR

4

Data Terminal Ready

COM

5

Common

DSR

6

Data Set Ready

RTS

7

Request to Send

CTS

8

Clear to Send

RI

9

Ring Indicator

The data sent is status information, in digital form, updated every two

seconds. Status is reported in the following order:

The concentration in ppm or percent