beautypg.com

6troubleshooting, Transporting and storing, Fig. 5 – Tanita 4521 User Manual

Page 7

background image

6

TROUBLESHOOTING

The following guide will help you to solve
common problems:

TRANSPORTING AND STORING

See Fig. 5 for visual assistance.

1. Unplug the adapter

from the outlet, then
unscrew it from the
electronic control
box and set the
adapter aside.

2. Slide the adjustable

rail to its widest setting.

3. Loosen the locking screw

and slide it off the black tabs.

4. Slowly lift the scale by the handle.

Avoid letting the rails bang together.

5. Wrap an elastic cord (64010) around each end of

the rails and hook the cord ends together.

6. Unscrew the scale cable from the control box and

firmly place the control panel box on the velcro tape
of the center column. Wrap the cable around the
center column above the control panel box and secure.

7. The scale is now ready for transportation or storage.

8. Remember to keep the adapter with the scale.

9. For additional convenience and protection of your

portable wheelchair scale, a scale storage bag is
available from your Tanita representative.

KILO

S

ON

OFF

Pounds

Kilos

www.ta

nita.co

m

Push “

ON” b

utton

and w

ait un

til zero

appea

rs befo

re

steppin

g on s

cale

WEIGHT

POUND

S

ZERO

Fig. 5

PROBLEM

Incorrect patient
weight readings

Patient weight readings
consistently too high
or too low

Display flashes “Lo-b”

SOLUTION

A. Platform is too low and hitting platform feet or

carpet. Increase platform height. See “Leveling
the Platform” on page 2.

B. Person, other than person being weighed,

is touching or leaning on scale handlebar

C. Weight of wheelchair or chair is not taken into

account. See “Operating Instructions” on page 4.

D. Platform is not level. See “Leveling the Platform”

on page 2.

Check scale calibration and calibrate if necessary.
See “Calibration” on page 5.

Recharge battery. See “Charging the Battery” on
page 3.