4. Calibration
The OME-A-822PGL/PGH is factory calibrated for optimum performance. Recalibration is
suggested for high vibration environments. The following items are required for
calibrating the OME-A-822PGL/PGH.
z
One 6 digit multimeter
z
One stable voltage source (4.9988V)
z
Diagnostic program : this program included with the
OME-A822PGL/PGH.
4.1 Description of Variable Resistors
There are seven variable resistors(VRs) on the OME-A-822PGL/PGH used for calibration,
they are described below.
VR Num. Description
VR1
A/D offset adjustment
VR2
A/D gain adjustment
VR3
D/A channel 0 gain adjustment
VR4
D/A channel 1 gain adjustment
VR5
D/A reference voltage adjustment
VR6
A/D unipolar offset adjustment
VR7
A/D programmable amplifier offset adjustment