Hh 504 / hh 505 calibration procedure, 00 w calibrator – Omega Engineering HH505 User Manual

Page 13

background image

1

HH 504 / HH 505 CALIBRATION PROCEDURE

Note: The following calibration procedure should perform only by qualified technicians

who have access to the items as following.

Equipment: The class of calibrator had better 10 times greater than the measured meter.

1. Turn off the meter then set the adjusted jumper to J1.

2. Insert the 383.00Ω calibrator to the T1 then turn on the meter, the main

display reading reads 1830.0 then press " ENTER " key.

3. The second display is stabilized (about 5sec.), then press " ENTER " key

when the meter auto power off then set the jumper back to J3 position.

4. Turn on the meter then Insert the 100.00Ω calibrator. The display reading

reads 0°C if the calibration procedure is right.

Pt385

Pt3926

-200℃

18.521Ω

16.996Ω

-100℃

60.256Ω

59.479Ω

0℃

100.000Ω

100.000Ω

100℃

138.505Ω

139.272Ω

200℃

175.856Ω

177.362Ω

300℃

212.052Ω

214.275Ω

400℃

247.092Ω

250.018Ω

500℃

280.977Ω

284.591Ω

600℃

313.708Ω

317.994Ω

700℃

345.280Ω

800℃

375.700Ω

383.00

W CALIBRATOR

+

RW1

RW1=RW2=RW3

RW=WIRERESISTANCE

G

+

_

_

RW2

383.00

W

G

RW3