Omega Engineering FLR1000 User Manual
Page 27

pg. 27 of 29
Symptom
Possible Cause
Method of Correction
Flow indication is not
linear and output too 
low at lower flows 
Attempting to measure
flows below the specified 
minimum for the unit 
 
Moisture or liquid in sensor 
 
 
Mechanical sensor assembly 
damaged 
 
Use higher flow rates
 
 
 
Remove moisture and allow 
sensor to dry out 
 
Unit must be returned for 
repair (see Section J) 
 
Flow reading accuracy
is outside the 
specification 
Gas has a specific gravity
other than Air at 20°C 
 
 
Moisture in sensor 
 
 
Mechanical sensor assembly 
damaged 
 
Pressure drop issue 
 
Calculate the flow reading
error / offset as explained in 
section C6 
 
Allow sensor to dry out. 
Ensure gas is dry. 
 
Unit must be returned for 
repair (see Section J) 
 
Correct for the pressure 
drop ( ΔP ) 
Flow indication is fixed
at one value
The maximum rated flow for
the unit is being exceeded. 
 
Output circuits may be 
damaged 
 
Reduce the flow
 
 
Unit must be returned for 
repair (see Section J) 
 
 
 