PART 7
APPROVALS INFORMATION
7.1 CE APPROVALS INFORMATION

This product conforms to the EMC directive 89/336/EEC amended by
93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.

Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and
laboratory.
Double Insulation
Pollution Degree 2
Dielectric withstand Test per 1 min

Power to Input/Output:

2300Vac (3250Vdc)

Power to Input/Output:

1500Vac (2120Vdc)

(Low Voltage dc Power Option*)

Power to Relays/SSR Output:

2300Vac (3250Vdc)

Ethernet to Inputs:

1500Vac (2120Vdc)

Isolated RS232 to Inputs:

500Vac (720Vdc)

Isolated Analog to Inputs:

500Vac (720Vdc)

Analog/Pulse to Inputs:

No Isolation

Measurement Category I

Category I are measurements performed on circuits not directly connected to the
Mains Supply (power). Maximum Line-to-Neutral working voltage is 50Vac/dc.
This unit should not be used in Measurement Categories II, III, IV.

Transients Overvoltage Surge (1.2 / 50uS pulse)

Input Power:

2500V

Input Power:

1500V

(Low Voltage dc Power Option*)

Ethernet:

1500V

Input/Output Signals:

500V

Note: *Units configured for external low power dc voltage, 12-36Vdc

EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement,
control and laboratory.

EMC Emissions Table 4, Class B of EN61326
EMC Immunity** Table 1 of EN61326

Note: **I/O signal and control lines require shielded cables and these cables must

be located on conductive cable trays or in conduits. Furthermore, the length
of these cables should not exceed 30 meters

Refer to the EMC and Safety installation considerations (Guidelines) of this
manual for additional information.

7.2

FCC
This device complies with Part 15, Subpart B, Class B of the FCC rules.